User Contributed MET/CAL PROCEDURE ============================================================================== INSTRUMENT: DANA 3860 PERFORMANCE CHECKS DATE: 15-Nov-88 AUTHOR: User Contributed REVISION: V 4.0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 17 NUMBER OF LINES: 37 CONFIGURATION: Fluke 5100B ============================================================================== STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- F 1.002 DISP Apply power and allow a 30 minute warmup. 1.003 DISP Set range switch to 200mV and short the input. 1.003 DISP Adjust the front panel offset for an indication of .000. 1.004 HEAD {DC VOLTAGE CHECKS} 1.005 5100 200 190mV 0.15% 2W 2.001 5100 200 -190mV 0.15% 2W 3.001 5100 2 1.9V 0.15% 2W 4.001 5100 2 -1.9V 0.15% 2W 5.001 5100 20 19V 0.15% 2W 6.001 5100 20 -19V 0.15% 2W 7.001 5100 200 190V 0.15% 2W 8.001 5100 200 -190V 0.15% 2W 9.001 5100 1000 700V 0.15% 2W 10.001 5100 1000 -700V 0.15% 2W 11.001 HEAD {RESISTANCE PERFORMANCE CHECKS} 11.002 DISP CONNECT TESTS LEADS TO OHMS AND COMMON TERMINALS. 11.003 RESF 200 100Z 0.45% 2W 12.001 RESF 2 1KZ 0.25% 2W 13.001 RESF 20 10KZ 0.25% 2W 14.001 RESF 200 100KZ 0.25% 2W 15.001 RESF 2 1MZ 0.45% 2W 16.001 RESF 20 10MZ 1.05% 2W 17.001 DISP PERFORMANCE CHECKS ARE NOW COMPLETE. 17.002 END